Abstract

In this paper, we present an investigation of the influence of the gas chromatographic separation system on the determination of polybrominated diphenyl ethers (PBDEs). Capillary columns, retention gaps and press-fit connectors, as well as different injection techniques have been evaluated with respect to yield and repeatability. The split/splitless injection has been optimized and compared to on-column injection, the septum equipped temperature programmable injector (SPI) and the programmable temperature vaporizing (PTV) injector. Furthermore, a comparison of the different operational modes of the PTV injector is presented. The results show that there are large variations in the yield of PBDEs depending on the column and the injection systems. Especially the high molecular weight BDE congeners can be subject to severe discrimination. Unfavorable conditions can lead to a complete loss of nona and deca substituted BDE congeners.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.