Abstract

Metal artifacts caused by hip arthroplasty stems limit the diagnostic value of computed tomography (CT) in the evaluation of periprosthetic fractures or implant loosening. The aim of this ex vivo study was to evaluate the influence of different scan parameters and metal artifact algorithms on image quality in the presence of hip stems. Nine femoral stems, 6 uncemented and 3 cemented, that had been implanted in subjects during their lifetimes were exarticulated and investigated after death and anatomical body donation. Twelve CT protocols consisting of single-energy (SE) and single-source consecutive dual-energy (DE) scans with and without an iterative metal artifact reduction algorithm (iMAR; Siemens Healthineers) and/or monoenergetic reconstructions were compared. Streak and blooming artifacts as well as subjective image quality were evaluated for each protocol. Metal artifact reduction with iMAR significantly reduced the streak artifacts in all investigated protocols (p = 0.001 to 0.01). The best subjective image quality was observed for the SE protocol with a tin filter and iMAR. The least streak artifacts were observed for monoenergetic reconstructions of 110, 160, and 190 keV with iMAR (standard deviation of the Hounsfield units: 151.1, 143.7, 144.4) as well as the SE protocol with a tin filter and iMAR (163.5). The smallest virtual growth was seen for the SE with a tin filter and without iMAR (4.40 mm) and the monoenergetic reconstruction of 190 keV without iMAR (4.67 mm). This study strongly suggests that metal artifact reduction algorithms (e.g., iMAR) should be used in clinical practice for imaging of the bone-implant interface of prostheses with either an uncemented or cemented femoral stem. Among the iMAR protocols, the SE protocol with 140 kV and a tin filter produced the best subjective image quality. Furthermore, this protocol and DE monoenergetic reconstructions of 160 and 190 keV with iMAR achieved the lowest levels of streak and blooming artifacts. Diagnostic Level III . See Instructions for Authors for a complete description of levels of evidence.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call