Abstract
An investigation was carried out to explore further the analytical capabilities of quadrupole and magnetic sector field inductively coupled plasma mass spectrometry (ICP-MS) in conjunction with ultrasonic nebulization (UN) for the reliable determination in human urine of nine potentially toxic and/or environmentally relevant trace elements, namely Cd, Co, Cr, Ni, Pb, Rb, Sb, Se and V. For the quantification of Cr, Ni and V magnetic sector field ICP-MS was used in the medium resolution mode (m/Δm=3000) to separate the signal of the analyte from severe polyatomic interferences caused by high concentrations of bulk elements (C, Ca and Cl). The detection power afforded by quadrupole ICP-MS, on the other hand, was adequate for the reliable determination of Cd, Co, Pb, Rb, Sb and Se. Three different sample pre-treatment steps, namely dilution of urine with high-purity water, UV irradiation and microwave digestion with concentrated HNO3 and H2O2, were investigated in detail to clarify two key issues, i.e., first, the performance, accuracy and reproducibility of the two ICP-MS systems under different matrix loads of the plasma and, second, the actual concentrations of trace elements in the urine sediment.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.