Abstract

ObjectiveWhen the compound muscle action potential (CMAP) is recorded in motor nerve conduction studies, the reference (E2) electrode can make a significant contribution to the CMAP. This study investigates the E2 recorded signal and its effect on CMAP measurements when E2 electrode is placed at different sites. MethodsThe CMAP was recorded using the active electrode on the muscle belly and 4 different E2 electrodes placed at distal and proximal sites. The signal recorded by each electrode was also measured using a reference electrode on the contralateral limb. Signals were recorded from the abductor pollicis brevis, abductor digiti minimi, tibialis anterior and biceps muscles. ResultsThe E2 recorded a smaller signal when it was placed near or off the proximal tendon or muscle origin. This affected CMAP latency, duration for tibialis anterior. Contrary to expectation, initial upward deflection was noted for E2 signal. ConclusionA proximal E2 position records a lower volume conducted signal and yields a CMAP that is more representative of the muscle over which the E1 (active) electrode is placed. SignificanceThe proposed ‘Proximal E2’ montage may be better suited to assess pathology.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.