Abstract

Objectives To compare HPLC methods with short and long elution times for HbA1c measurement in blood. Methods Comparison between G7-Tosoh (1.2 min); Bio-Rad-Variant-II-Turbo (1.3 min) and Arkray-HA-8160 (2.9 min). Results Passing–Bablok regression equations were: Y = 0.17 + 0.96 X; Y = − 0.39 + 1.01 X; Y = − 0.40 + 1.0 X and the means of the differences using Bland–Altman Plot were 0.02; − 0.34; 0.32 for HA-8160/G7-Tosoh, HA-8160/Variant-II-Turbo and G7-Tosoh/Variant-II-Turbo, respectively. Conclusions Faster elution methods had no problems on reproducibility of results obtained by slower elution methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.