Abstract

Squeaky ceramic-on-ceramic (COC) total hips have received much recent publicity, and implant design may be implicated. We reviewed 270 consecutive COC total hip arthroplasties in 233 patients comparing 4 implant combinations representing 4 manufacturers. A cohort (n = 45) of Stryker Trident acetabular cups paired with Stryker Accolade femoral stems showed a dramatically higher incidence of “problem squeaking”—defined as always audible to others and occurring at least once per week—with a 35.6% incidence of squeaking and 11.1% incidence of problem squeaking. The 3 non-Stryker designs (n = 225) revealed 3.6% squeaking ( P < .0001) and 0.44% problem squeaking ( P = .006). The Stryker system has a unique design and metallurgy. Our results suggest that although the genesis of squeaking in COC total hips is multifactorial, prosthetic design plays a key role.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.