Abstract

AbstractThe objective of a continuing study by the Belgian Particle Technology Group involves difficulties in practical size determinations. This part describes the behaviour of silicas with different porosities during sizing by electrical sensing zone and laser diffraction methods. Anisotropy and porosity were identified to be important particle characteristics in understanding the differences between the two methods. Especially large pore diameters and pore volumes were found to be responsible for shifts in size distribution of 50 to 100%. The use of optical values and optical models was shown to influence these shifts considerably. In the case of spherical silica particles with moderate porosity, no significant differences could be found between the two sizing methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.