Abstract
Reflectance anisotropy spectroscopy (RAS) can be used to monitor (reactive) ion etching (RIE) of semiconductor samples. We present results on the influence of the Cl2 content of the plasma gas on the RAS spectra during reactive ion etching. In a first step GaAs samples have been used and the RAS spectra are compared to results of secondary ion mass spectrometry (SIMS) on sample surfaces and depth profiles. In a second step a III–V semiconductor multilayer system has been investigated using the time-evolution of the average reflected intensity as an indication for the etch rate. In both cases usually even a high amount of Cl2 does not disturb the surface-sensitivity of the RAS signal.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.