Abstract

A change in the position of a prosthetic acetabular component between two different radiographic examinations indicates loosening, and may be observed as tilting or migration of the socket. To determine the apparent changes in socket position caused by improper positioning of the pelvis, a full-scale model of a pelvis with attached prosthetic sockets was examined radiographically in different positions. The projected alignment of the Charnley socket indicator wire against the inter-tuberosity line was markedly influenced by the positioning of the pelvis. An alternative way of measuring the alignment is to use the long axis of the projected ellipse of the outer circumferential groove in the socket polyethylene. The thus defined socket alignment was not influenced by the positioning of the pelvis within the investigated range. No apparent socket migration was recorded within the range of pelvic rotation and inclination studied.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.