Abstract
Objective. To examine the influence of the endotracheal tube (ETT) on respiratory reactance (Xrs) measured with the forced oscillation technique (FOT) and develop a correction method for it. Approach. In a bench study, the reactance of ETTs (Xtube) with different dimensions was measured on a breathing test lung in various respiratory settings. Main results. Xtube can be accurately predicted by a fitted formula, with an R2 of 0.97, with negligible effects due to changes in respiratory pattern and lung volume. Significance. The developed formula offers the ability to measure ETT-independent Xrs values of patients, improving the potential of FOT for lung function testing in mechanically ventilated newborns.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.