Abstract
This in vitro study was conducted to investigate the repeatability of the implant stability quotients (ISQ) measured with multipegs after numerous sterilizations and to detect the exact time when the readings start to deviate. Multipegs were sterilized with 3 different methods (autoclaved, autoclaved + ultrasonic cleaner, chemical disinfection + autoclaved) and grouped according to the method applied. All specimens were put into the autoclave with sealed packages every time they were sterilized. Each specimen was sterilized 50 times according to the technique described in its group after an ISQ measurement was performed. Results of the 2-way analysis of variance showed that neither the sterilization method nor the cycles, nor their interaction, were statistically significant. A multipeg may be reused multiple times after sterilization procedures and may be more cost-effective than a disposable smartpeg for checking implant stability after confirming these results in further investigations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.