Abstract

Clinical studies show that non-invasive respiratory support by continuous positive airway pressure (CPAP) affects gas conditioning in the upper airways, especially in the presence of mouth leaks. Using a new bench model of neonatal CPAP, we investigated the influence of mouth opening on oropharyngeal temperature and humidity.The model features the insertion of a heated humidifier between an active model lung and an oropharyngeal head model to simulate the recurrent expiration of heated, humidified air. During unsupported breathing, physiological temperature and humidity were attained inside the model oropharynx, and mouth opening had no significant effect on oropharyngeal temperature and humidity. During binasal CPAP, the impact of mouth opening was investigated using three different scenarios: no conditioning in the CPAP circuit, heating only, and heated humidification. Mouth opening had a strong negative impact on oropharyngeal humidification in all tested scenarios, but heated humidification in the CPAP circuit maintained clinically acceptable humidity levels regardless of closed or open mouths.The model can be used to test new equipment for use with CPAP, and to investigate the effects of other methods of non-invasive respiratory support on gas conditioning in the presence of leaks.