Abstract
The influence of monitor brightness and room illumination on soft-copy diagnosis by both cathode-ray tube (CRT) monitor and liquid crystal display (LCD) was evaluated and compared using a contrast-detail phantom. Nine observers (7 radiologists and 2 radiological technicians) interpreted six types of electronically generated contrast-detail phantom images using a 21-inch CRT (2,048x2,560) and a 21-inch LCD (2,048x2,560) under 6 kinds of viewing conditions, i.e. monitor brightness of 330 cd/m2 or 450 cd/m2, and room illumination of 20, 100 or 420 lux at the center of the display. Observers were requested to determine the visible borderline of the objects. Between 330 cd/m2 and 450 cd/m2, no significant difference in the visible area was found under any of the three lighting conditions. However, in two low-contrast phantom images, the visible area on the LCD was significantly larger than that on the CRT, independent of both monitor brightness and room illumination. (p<0.05). The effect of room illumination was not significant, suggesting that the use of LCD at high room illumination is acceptable.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.