Abstract

A portable breath carbon monoxide (CO) monitor has a high cross-sensitivity to hydrogen (H2). This study examined the influences of H2 after consuming milk on the detected CO values using three types of portable CO monitors. Exhaled breath from seven participants (four healthy nonsmokers and three smokers with otherwise unknown comorbidities) was collected in sampling bags. The participants then consumed 200 mL of milk, and the exhaled breath of each was collected in separate bags every 30 minutes until 9 hours later. CO and H2 in the bag were measured using a gas chromatograph as a reference analyzer, and CO was also measured using three types of portable CO monitors. After consuming milk, H2 levels were significantly higher, and CO levels were not significantly elevated as measured by the reference analyzer. However, CO levels in monitors A and B were significantly elevated, even though participants did not smoke. The H2 levels in the reference analyzer significantly increased and reached a maximum 4.5 hours after consuming milk. The difference in CO levels between the reference analyzer and each monitor increased significantly after 5 or 5.5 hours. This study suggested that the breath CO monitors with a cross-sensitivity to H2 responded to H2 as CO in the exhaled gas and measured higher than actual values after milk consumption. The extent of the influence of H2 differed depending on the type of CO monitor. It is necessary to consider milk consumption when assessing the smoking status of people using portable CO monitors.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call