Abstract

This study evaluates the accuracy of demodulated sound measurements using a condenser microphone in the near field of a parametric loudspeaker system. Microphones with different sensitivities placed at incidence angles of 0° and 90° were used to measure demodulation frequency components without special acoustic filters. The measured components were compared with theoretical predictions. The results show that the measured sound pressure using microphones placed at 0° was up to several tens of decibels larger than the theoretical predictions and significantly inaccurate in the near field. This was due to the nonlinear response of the microphone, which had high sensitivity at primary sound frequencies, inducing spurious signals. This result suggests that using a microphone with low sensitivity at primary sound frequencies placed at an appropriate angle that reduces sensitivity improves parametric sound measurement accuracy.