Abstract
Continuous spinal analgesia (CSA) offers considerable pain relief, and has been used in various procedures such as for surgery and cancer pain control. In Japan, portable and disposable infusion kits are increasingly employed for continuous epidural analgesia and CSA. In CSA, the use of a microcatheter is expected to reduce the incidence of spinal headache previously encountered with larger catheters. However, the flow rate of disposable infusion kits is reduced when used in conjunction with a microcatheter. This study aimed to investigate the influence of catheter length on the flow rate of two different devices: 20- or 91-cm 28-G microcatheters connected to balloon- or syringe-type infusion pumps were examined (50 mL, 1 mL h(-1)). There were four groups each of 10 experiments: Group A: balloon-type infuser, 91 cm catheter; Group B: balloon-type infuser, 20 cm catheter; Group C: syringe-type infuser, 91 cm catheter; Group D: syringe-type infuser, 20 cm catheter. The mean flow rate in Group A was significantly less than that in Group B and the mean flow rate in Group C was significantly less than Group D (P < 0.05). These results indicated that the use of a shorter microcatheter achieves a better flow rate during CSA.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.