Abstract

Influence of the measured objects parameters on coordinate measuring machine (CMM) touch trigger probe inaccuracy of probing is investigated. Factors such as material stiffness of measured workpiece, its surface shape and roughness are taken into consideration. The measuring method out off CMM, based on a rotary table and a reference axis is applied for tests. The concept is based on measurements of the distance between reference and triggering points in various directions. The reference points are established by the rotation of a precise rotary table. The statistical analysis of variance and regression is applied for data analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.