Abstract

Patient-specific finite element models (PSFEM) are becoming more and more used. Different methods for assigning their material properties were studied on PSFEMs of 9 tibias along with the minimal required length of the CT acquisition window. Material properties are generally attributed to the PSFEM using relationships linking the grayscale of CT scans to the elasticity moduli. Using cortical-specific and trabecular-specific relationships or a generic one, did not result in significant differences. However, the use of homogeneous elastic moduli in the cortical and trabecular regions led to considerable differences. The result highlight that the PSFEM must comprise at least 40% of the tibia to ensure consistent results in the proximal 20%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.