Abstract

In this work, the influence of the sample gas flow rate and the ionization region volume of an ion mobility spectrometer (IMS) used as a detector in gas chromatography (GC) on GC-IMS peak shape has been investigated. Therefore, a drift tube IMS with a field-switching ion shutter, a defined ionization region volume and an ultra-violet radiation source was used. To identify the influence of the sample gas flow rate entering the ionization region (equals the GC carrier gas flow rate if no further make-up gas is used) and the ionization region volume on peak broadening and signal intensity, different sample volumes as they would elute from a GC were tested at a variety of sample gas flow rates at a given ionization region volume. The results clearly show that for low sample gas flow rates a depletion of sample molecules in the ionization region leads to a significant decrease in effective detector volume but also to reduced signal intensities. Therefore, for optimal performance of a GC-IMS, the optimal operating point of the GC should match the flow range, where the IMS provides the best compromise between signal-to-noise ratio and peak broadening.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call