Abstract

The energy distributions of ions produced in a spark source mass spectrometer were recorded for different trace element ions and for the total ion current. Both electrodes of pure doped graphite, and with 10% potassium salts added, were examined. The results show that the measured energy of an ion is mainly determined by the m/z ratio and that the energy of the total ion current depends on the sample composition. It was also clear that if the accelerating voltage is set to obtain maximum transmission of the total ion current, the fraction of the heavy elements that is discriminated at the β-slit depends on the matrix composition if the pass band of the instrument is narrow. These effects influence considerably the accuracy in the analysis of variable samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.