Abstract
The effect of ion lens voltage settings and plasma conditions on nonspectroscopic interferences has been studied. The change in ion count rate for 7Li, 10B, 45Sc, and 89Y in the presence of Na, K, Sr, and Cs under a variety of experimental conditions is reported. The ion lens voltage setting did not affect the degree of nonspectroscopic interferences with the use of the Einzel ion optic system. Suppression effects were increased when the power was increased and the sampling depth decreased. Changes in nebulizer pressure resulted in interference trends, which varied with the concomitant element being studied. Sodium caused a decrease in interference effects with increasing nebulizer pressure, whereas Cs caused an increase under the same experimental conditions. Suppression of analyte ion count rate increased with increasing sample solution uptake rate and increasing atomic mass of concomitant elements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.