Abstract

Background/purposeDental implants are a mainstream solution for missing teeth. For the improvement of dental implant surface treatment and design, short dental implants have become an alternative to various complex bone augmentation procedures, especially those performed at the posterior region of both the maxilla and mandible. The objective of this study was to evaluate the effect of various insertion methods on the primary stability of short dental implants. Materials and methodsCommercial dental implants were inserted into artificial mandibular bone specimens using various insertion methods (equicrestal position, subcrestal position 1.5 mm, and lateral cortical anchorage) in accordance with an implant surgical guide. Insertion torque value (ITV) curves were recorded while implant procedures were performed. Both maximum ITVs (MITVs) and final ITVs (FITVs) were evaluated. Subsequently, Periotest values (PTVs) and implant stability quotients (ISQs) were measured for all specimens. A Kruskal–Wallis test was conducted to analyze the results for four primary stability parameters, and the Dunn test was used for a post hoc pairwise comparison when a difference was identified. ResultsFor all groups, their mean MITVs ranged from 33.6 to 59.4 N cm, whereas their mean FITVs ranged from 17.5 to 43.5 N cm. Insertion torque value, ISQ, and PTV decreased significantly when implants were inserted into subcrestal positions. When implants were inserted in the lateral bicortical position, the four aforementioned parameters yielded greater values. ConclusionWhen 6-mm short implants were inserted in a lateral cortical anchorage position, high primary stability was yielded.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call