Abstract

BackgroundRegadenoson is a novel pharmacological stress agent whose disposition during hemodialysis is not known. The purpose of this study was to determine the clearance of regadenoson under varying dialytic conditions using an in vitro hemodialysis model. Methods and ResultsWhole human blood was used to analyze the effect of hemodialysis on the clearance of regadenoson. Regadenoson transmembrane clearance (CLD) was assessed for both a standard permeability and a high permeability polysulfone hemodialyzer with blood/dialysate flow rates of 300/600 and 400/800 mL/min. A two-tailed, paired Student’s t test was used to compare regadenoson CLD between hemodialyzer types and flow rates. The mean ± SD regadenoson CLD values ranged between 62.5 ± 11.8 and 89.1 ± 24.0 mL/min for all dialytic conditions. There was no significant difference in regadenoson CLD between hemodialyzer types and flow rates (p > .05). ConclusionsHemodialysis enhances the clearance of regadenoson independent of hemodialyzer permeability and blood/dialysate flow rate. This clearance is modest relative to total body clearance and is unlikely to produce a clinically significant outcome.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.