Abstract

Five methods for fat extraction from powdered milk infant formulas were tested for reproducibility and accuracy of the peroxide value determination. n-Hexane, n-hexane after mortar grinding in the presence of some water, n-hexane/methanol, dichloromethane/methanol/petroleum ether and n-hexane/isopropanol were used for the extraction. Among all the tested methods, n-hexane/isopropanol 3:1 mixture gave quantitative yields in fat extraction from reconstituted milk. This method combined with the classic iodometric titration gave the best result in terms of reproducibility, detecting peroxide values down to 1.1 (n=5, RSD=10%) and 1.3meqO2/kg for 4 and 10g of milk respectively. All these features, the rapidity, low cost, easiness, as well as the relatively low amount of sample employed for the analysis with no need of any special instrumentation, make it suitable as a routine method for the evaluation of the quality of such irreplaceable dietetic products as milk infant formulas are.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.