Abstract
Spectroscopic and electrothermal conditions for the determination of antimony in seawater using a transversely heated graphite furnace with Zeeman-effect background correction have been optimized with the use of an a priori calculation of the detection limit. The lowest limit of detection was obtained with a 2 nm spectral curvatures bandwidth and the use of an electrodeless discharge lamp; however, these experimental conditions resulted in strong premature curvature of calibration curves. Pd(NO(3))(2) can be recommended as a chemical modifier because seawater interference effects are minimized and pretreatment curves up to 1500 degrees C can be used permitting the removal of the major part of the saline matrix before atomization. Under optimized spectroscopic and electrothermal conditions the obtained limit of detection of Sb in seawater was about 0.4 microg L(-1).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.