Abstract

Molecular mobility is an important measure in biological functionality, as molecules have to diffuse to meet and interact and perform actions. Measurement of mobility requires specific tools such as fluorescence correlation spectroscopy (FCS). Especially, combination with superresolution stimulated emission depletion microscopy (STED-FCS), whether in a point- or beam-scanning mode, has proven valuable for determination of anomalous diffusion. STED-FCS however relies on an accurate calibration of the effective observation spot formed for different laser powers of the additional STED laser. This poster article highlights the need for calibration measurements and outlines that rather simple procedures involving acetone cover-glass surface cleaning only, instead of piranha cover-glass surface cleaning, and point instead of more complex scanning STED-FCS are sufficient for calibration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.