Abstract

Introduction: The aim of this study was to investigate the role of the cross-sectional design surface in improving the cyclic fatigue resistance by comparing two nickel–titanium (NiTi) rotary instruments with the same heat-treatment and grinding procedure. Materials and Methods: A total of 40 new NiTi instruments (20 S-One 25.06 taper and 20 AF Blue S4 25.06 taper) were tested in the present study. Both instruments were rotated at 300 rpm and with a torque setting of 2 Ncm) using an endodontic motor in the same artificial canal (90° angle of curvature and 3 mm radius). A Student's t-test was performed to determinate the differences in terms of time to fracture (TtF) and the length of the fractured portion between the two different instruments. Significance was set at P Results: The differences in terms of TtF for cyclic fatigue resistance between the two files were statistically significant (P 0.05). Scanning electron microscopy images showed patterns of cyclic fatigue striations. Conclusion: The results of this study confirm and enhance the role of cross-section in increasing the cyclic fatigue resistance of NiTi rotary instruments.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call