Abstract

Germanium crystals have a dead layer that causes a decrease in efficiency since the layer is not useful for detection but strongly attenuates photons. The thickness of this inactive layer is not well known due to the existence of a transition zone where photons are increasingly absorbed. Therefore, using data provided by manufacturers in the detector simulation model, some strong discrepancies appear by changing the dead layer. Investigations into the Ge detector response functions for gamma rays have been conducted using straightforward physical mechanisms implemented by Monte Carlo simulations. The detector response function feature's most probable interaction mechanisms are described. The Monte Carlo method is applied to simulate the calibration of a HPGe detector in order to determine the total inactive germanium layer thickness and the active volume that is needed in order to study the response function for both types of detectors. Results indicated a strong impact of dead layer variations on the response function of the simulated detectors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.