Abstract

Information about blood arterial oxygen saturation (SpO2) is crucial in critical care settings or home health monitoring during the COVID-19 pandemic. Also, we need to identify the factors that affect the SpO2 measurement. In this paper, the effect of compression of the cuff during noninvasive blood pressure (NIBP) measurement on the SpO2 results was investigated. A custom-made system was used for simultaneous measurement of NIBP and SpO2. The study was conducted on 213 subjects aged between 21 and 93, with asystolic blood pressure of (94 to 194) mmHg, diastolic blood pressure of (52–98) mmHg, and 994 NIBP readings were used for the analysis. During the NIBP measurement, momentary changes in SpO2 can reach ±17% and are in most cases positive (mean 2.9%). The change was not correlated with sex, age, height, body weight, BMI, HR and blood pressure. The obtained results show that frequent NIBP measurements may lead to wrong conclusions about SpO2. In our study, pressure measurements mainly caused the increase of blood oxygenation level.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.