Abstract

Several clinical reports pointed out that the monoclinic transformation of crystalline structure of zirconia femoral head with clinical service time can cause the changes of surface roughness and mechanical properties. To elucidate the relationship between these surface roughness change induced by phase transformation and UHMWPE socket wear, hip simulator wear tests were performed on total hip prostheses (THP) up to 6 million cycles running. Four groups of different surface monoclinic phase fraction of Kyocera zirconia heads were prepared by aging treatment prior to the wear tests. The surface roughness/ wear rate of zirconia heads with 0, 10, 50 and 80mol% monoclinic phase fraction were 0.009/ 3.8, 0.014/ 3.9, 0.012/ 3.7 and 0.010µmRa/ 3.1mg/million cycles, respectively. These results exhibit that the surface roughness change induced by phase transformation does not significantly influence extent of the generation of UHMWPE debris in Kyocera zirconia /UHMWPE THP system. In contrast, the surface roughness exhibited complex relation with the monoclinic phase fraction and therefore the morphological analysis was performed on the zirconia articulating surfaces

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.