Abstract
Thermoluminescence dosemeters are widely used to monitor personal doses. For these low dose range applications, it is important to determine the detection limit L(D) and the determination limit L(Q) of the dosimetric system. The influence of background exposure on these limits for LiF:Mg,Cu,P(GR-200A) based TL dosimetry was investigated. Both the conventional analysis and the glow curve analysis methods were used to determinate these limits. The detection limit L(D) was compared with the recording level and the investigation level. A systematic error can occur in the occupational dose evaluation when the detection limit L(D) is more than the recording level. It was found that the L(D) of the dosimetric system-based LiF:Mg,Cu,P(GR-200A) was less than the recording level for exposure time tau > or = 10 days considering an annual dose limit of 1 mSv for the public recommended in ICRP Publication 60.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.