Abstract

Abstract An ultrasonic-assisted bone instrument can be used for the dissection of bone in spinal surgery. During surgery, the tip of such a bone instrument can touch spinal dura mater. Especially during critical re-operations, high forces can be unintentionally applied on spinal dura mater. To prevent dural tears, the influence of application parameters on the tear force shall be analyzed. Collagen foil is used as a substitute material for spinal dura mater because of its similar mechanical properties and structure. Sponge cloth is placed below collagen foil to imitate cerebrospinal fluid and nerve tissue. A 3-axis CNC-machine is used to automate the movement of the bone instrument. In a full factorial experiment the influence of amplitude, shape of movement and velocity of movement on the tear force is analyzed. Amplitude has the strongest effect because of the increase in input energy. Velocity also has a significant influence, probably because of the strain rate increasing with velocity. Thus, a surgeon should mainly reduce the amplitude and furthermore the velocity of movement to generate higher safety in critical areas.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.