Abstract

Studies of haemocompatibility of AISI 316LVM steel point to the need for nickel elimination from the surface and replacing it with other elements showing higher biotolerance. Such layers include titanium, carbon or silicon coatings. Therefore, the authors attempted to evaluate some selected physicochemical properties of TiO2 layers, grown by atomic layer deposition (ALD) method, on the surface of 316LVM steel at variable process temperature. ALD temperature has a major role in the final quality of the surface layer grown with the use of such method, regardless of the type of the base. It was observed that the growth of temperature had an adverse influence on corrosive resistance in the artificial plasma environment and contributed to formation of a double (porous) layer showing decreased tightness. Further on, assessment of the coating adhesion to the base showed that too low process temperature T=100°C had an adverse effect on mechanical properties, resulting in substantially reduced critical force. On the other hand, the performed surface wettability tests showed no influence of ALD temperature in the obtained angle values.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.