Abstract
Monitoring of intracranial pressure (ICP) provides a life-saving diagnostic tool. We present a battery-free pressure sensing system for minimally invasive ICP monitoring. It comprises a cranially concealed wireless pressure sensor and on- and off-body external units. The sensor is based on a piezoresistive element and is inductively powered through an on-body unit. It is also equipped with a far-field antenna, which conveys the pressure data to the off-body unit. We present the electromagnetic modeling of the system and report results from experiments carried out in a setting, which mimics the biological operation environment. The simulation and measurement results demonstrate the wireless pressure monitoring at a distance up to one meter over the pressure range from −3 to 33 mmHg and assess the impact of an imperfectly aligned inductive link on the operation of the system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.