Abstract

Spatially resolved, radical excitation temperature distributions experienced by analyte species injected into the axial channel of a toroidally shaped, inductively coupled argon plasma are presented. Typical axial temperatures experienced by the thermometric species (FeI) ranged from 6500 K to 5500 K in the analytical zone of 15–25 mm above the load coil. A comparison of temperatures calculated with different sets of transition probabilities is also given.