Abstract

Pulmonary vein isolation in a dual-chamber pacemaker patient using the pulmonary vein ablation catheter (PVAC) system resulted in perpetual induction of ventricular tachycardia (VT) during radio frequency energy application. Induction of VT was abolished by programming the PVAC-system to a pure bipolar ablation mode. Patients with implanted devices should be closely monitored when using the PVAC system in unipolar modes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.