Abstract

Inducible displacement CT compares 2 CTs acquired in series but with alternated rotation of the femur. This provides visual and quantitative clues as to the mechanical situation, i.e., loosening, of a total hip arthroplasty. We report the accuracy of this method as well as the experience of integrating it into a clinical workflow. This was a retrospective single centre study of 72 cases of suspected aseptic loosening were the surgeon after reviewing a standard plain radiograph saw a need for more information. The displacement CT and plain radiograph were compared either to intraoperative findings or a 1-3 year follow up questionnaire for patients that did not have revision surgery. Patients reporting degradation in status since the time of the displacement CT were called for a follow up plain radiograph. Sensitivity and specificity were assessed, and user experience gathered. Of 72 enrolled patients 15 were lost to follow up. Of the remaining 57, 17 were judged by in-traoperative findings or follow-up to have had loose implants. For plain radiography the sensitivity and specificity were 59% (95% CI 35-82) and 85% (74-96). For displacement CT the corresponding values were 77% (56-97), and 100% (100-100) respectively. The tool was adaptable to clinical routine. Displacement CT with alternated rotations of the femur is a viable option to improve the diagnostic process for identifying aseptic loosening in a total hip arthroplasty.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.