Abstract

We report extensive cyclic voltammetric evaluation of gold submicrometer edge band electrodes with different widths, from 25.3 to 143.5 nm. Fabrication of these band electrodes from layers of gold and silicon nitride using conventional microfabrication techniques, was reported in the preceding companion paper. Gold film thicknesses and, thus, the electrode widths were measured by surface profilometry, X-ray interferometry, and atomic force microscopy. The electrodes exhibit typical sigmoidal-shaped cyclic voltammograms at slow scan rates (<1 V/s). The limiting currents deviate from theoretical values as electrode width decreases. At larger widths, experimental values are in better agreement with theory, consistent with trends reported in the literature. Maximum current is much higher than expected for all widths at fast scan rates (> 25 V/s). Capacitance of the edge band electrodes is high (187−0.206 mF/cm2 for 71.2-nm edge band electrodes for scan rates of 0.010−204 V/s) and scan rate dependent. Values approach those of macroelectrodes at fast scan rates, indicating imperfect seals at the gold/silicon nitride, gold/Cr/glass, or gold/epoxy interfaces.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.