Abstract
Pedicle screws are used for the treatment of a wide variety of spinal pathologies. A good screw holding power in bone is required for treatment success, but has so far not been predictable computationally. The goal of this study was to develop an automated tool able to predict patient-specific screw fixation strength through finite element simulation. We compared the simulation results with results from biomechanical pull-out tests performed on animal lumbar specimens. Experimental and simulation pull-out strengths were highly correlated and the mean error was 20.25%. The fixation strength was also associated to great extent with pull-out stiffness and strain energy, as well as the screw size and mean vertebral density.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Computer Methods in Biomechanics and Biomedical Engineering
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.