Abstract

We assessed the repeatability and agreement of ganglion cell complex (GCC) in the macular area and the peripapillary retinal nerve fiber layer (ppRNFL) with individual and combined macula and disc scans. The macular GCC and ppRNFL thicknesses from 34 control eyes and 43 eyes with glaucoma were measured with the Canon Optical Coherence Tomography (OCT) HS-100. Two repeated measurements were performed with both scan modes. The repeatability limit (Rlim) and agreement analysis were performed. The individual scan showed better repeatability than the combined scan in both groups. However, the differences in the Rlim for the GCC in most sectors were lower than 3 μm (axial resolution of the OCT), and this was larger than 3 μm for most of the ppRNFL sectors. The mean differences in the thickness between both scan modes for the GCC and ppRNFL measurements were less than 3 and 6 μm, respectively. The interval of the limits of agreement was about 10 μm in some sectors for the GCC, and about 40 and 60 μm in some sectors in controls and glaucoma eyes, respectively. Both scan modes showed good repeatability in both groups. The agreement results suggest that the scan modes cannot be used interchangeably.