Abstract

The development of a simple ion chromatographic method for the analysis of citrate in pharmaceutical matrices is described utilizing the technique of indirect photometric detection. A polymeric anion exchange column was used with a mobile phase consisting of an aqueous solution of trimesic acid (pH 10) and indirect detection at 280 nm. The developed method was precise and accurate with percent recoveries (±%RSD) of 99.3 ± 0.97% and 98.2 ± 0.25% at 6 and 12 μg of citrate injected. Linear detector response was found over the range 1–12 μg of citrate injected with a limit of quantitation of 0.26 μg of citrate injected. The method is specific for tricarboxylic acids as most other ions were not retained by the column. Youden's robustness test, involving seven selected operating variables, showed that the method is fairly robust. A system suitability test was also performed. Many commercial USP and nondashUSP liquid and tablet formulations were assayed for citrate by the developed method and the results were found to be comparable with those obtained by the compendial method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.