In the present communication, it is shown that a laser scalpel operating at the wavelength of 0.97 μm with a fiber tip having a strongly absorbing coating proves to be efficient for making contact incisions. At 6 W laser pump radiation, the temperature of silica glass on the cutting fiber tip during cutting is close to the silica glass softening point. The high fiber tip temperature allows bloodless resection with the cut depth of more than 1 mm and with high tissue cutting speed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.