Abstract
A novel method was developed for the indirect determination of chlorophenols by AAS after extraction of their ionic associates involving the dipyridylocopper(II) (CuDP) or phenanthrolinocopper(II) (CuPH) complex into nitrobenzene, o-dichlorbenzene or methyl isobutyl ketone. Nitrobenzene appeared to be particularly suitable for the extraction of the CuDP associate, methyl isobutyl ketone, for the extraction of the CuPH associate. Thee method is applicable to chlorophenol concentrations of 10-6 to 10-7 mol l-1. Using CuDP after a single complete extraction into 10 ml of nitrobenzene, the lowest determinable concentration giving rise to absorbance A = 0.010 is 0.3 mg of pentachlorophenol in a litre; using CuPH after a single extraction into 10 ml of methyl isobutyl ketone, it is 0.2 mg of chlorophenol in a litre.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Collection of Czechoslovak Chemical Communications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.