Abstract
End-channel indirect amperometry is based on the principle of Kohlrausch regulating function (KRF). A dilute electroactive ionic species is added to the background electrolyte as a continuously eluting electrophore, which is used as probe. The probe concentration variation with the ω value of KRF in the sampling zone was described schematically in this report. Either cathodic or anodic electroosmotic flow (EOF) rates were monitored in microchip. There was no significant difference between the values of EOF rates measured by present method and current-monitoring method. Detection of electroactive and nonelectroactive analytes can also be accomplished by indirect amperometric method. Hence, the effective mobility of analytes can be accurately calculated. And the response mechanism of nonelectroactive analytes K +, Na + and Li + in the indirect method was speculated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.