Abstract

New developments on Rogowski coil are presented due to the evolution of temporal shortening of the current pulses. Recently, to study biophysical stresses by electromagnetic pulses flat transmission lines are used. As a result, linear Rogowski coils for electrical current diagnostics have also been developed and applied. An electrostatic shield with a slim slit is used to protect the Rogowski coil. The presence of the slit is necessary to lead the input current around the coil. The response of the Rogowski coil on high frequency signals is influenced, not only by all inductances, the capacitance, the number of turns of the coil, the load resistance, but also by the small size of the slim slit operated on the metal shield. The new theory is able to justify the response of the device due to sub-ns pulses, ~100ps, on slit width. By the theory, the attenuation factor is dependent on the slit size. This behaviour was experimentally verified obtaining a value ranging from 15.6A/V to 24.7A/V. The longest measurable pulse was approximately 250ns. The device is very interesting owing to the radiation of the 5G communication mode.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.