Abstract

The purpose of this study was to determine the protection factor of the Lepestok-200 filtering facepiece respirator by conducting a standard quantitative fit test on a panel of 25 representative adults (14 males and 11 females) using the TSI Incorporated PortaCount PlusTM quantitative fit-testing system. Each subject was tested four times. In the total of 100 tests, 95% of the overall fit factors were greater than 3, more than 80% of the overall fit factors were greater than 14, approximately 50% were greater than 86, and 20% were greater than 200. The pass-fail performance of the respirator was similar for each of the six exercises in the test series: (1) normal breathing, (2) deep breathing, (3) moving the head side to side, (4) moving the head up and down, (5) reading a passage of text out loud, and (6) normal breathing, indicating that the respirator performs equally well for each type of exercise. A significant and sustained improvement in fit factor was observed after the initial test, indicating that the subjects benefited from the knowledge gained in the first of the four quantitative fit tests. In the 75 tests conducted after the initial test for each individual, 95% of the overall fit factors were greater than 6, more than 80% of the overall fit factors were greater than 23, and 50% were greater than 116, and 20% were greater than 200. Thus, the initial learning experienced doubled the fit factor for subsequent tests. In addition, there is an indication that the Lepestok-200 may perform better on wearers with wider faces than on individuals with narrower faces. The results of this study demonstrate the effectiveness of the Lepestok-200 respirator and reinforce the general conclusion that quantitative fit-testing can make an important contribution to ensuring that proper protection factors are achieved for workers.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call