Abstract

The checking of monitor unit calculations is recognized as a vital component of quality assurance in radiotherapy. Using straightforward but detailed computer-based verification calculations it is possible to achieve a precision of 1% when compared with a three-dimensional (3D) treatment planning system monitor unit calculation. The method is sufficiently sensitive to identify significant errors and is consistent with current recommendations on the magnitude of uncertainties in clinical dosimetry. Moreover, the approach is accurate in the sense of being highly consistent with the validated 3D treatment planning system's calculations. PACS number(s): 87.53.–j, 87.52.–g

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.