Abstract
Incremental flip angle (IFA) snapshot fast low angle shot (FLASH) is a new modification of inversion recovery snapshot FLASH MR imaging. The method changes the flip angle incrementally from low to high during data acquisition and was applied in the evaluation of 16 focal hepatic lesions in 10 patients. Sequence comparisons were performed with a fixed flip angle inversion recovery snapshot FLASH sequence (standard), a T1- and T2-weighted spin-echo (SE) sequence, and a T1-weighted breath-hold FLASH sequence. Whereas snapshot FLASH images in both pulse sequences were free from physiological motion artifacts, SE and FLASH images showed respiratory artifacts in some patients. Quantitative analysis of IFA snapshot FLASH images at low hepatic and low lesion signal revealed both superior lesion-liver signal-difference-to-noise ratio (SD/N) and superior contrast compared with standard snapshot FLASH without additional artifacts. Unless motion artifacts were evident, SE and FLASH images showed a higher anatomic resolution but lower SD/N and lower contrast than IFA snapshot images. Because of its superior SD/N and contrast, IFA snapshot FLASH will likely widen the application of fast MR imaging techniques.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.