Abstract
This work demonstrates sampling of singly charged particles up to 200nm in diameter at atmospheric pressure into vacuum and trapping large numbers (>106) at a point in front of the end cap electrode of a linear quadrupole ion guide/trap for on-demand injection into the acceleration region of a time-of-flight mass spectrometer in a well-collimated ion packet. This procedure was shown to yield trapping efficiencies that ranged from 4 to 5% for 10nm diameter urea particles (∼400kDa) to 1% for 200nm urea particles (∼3×109Da). Analysis of the inlet optimization procedure suggests that the inlet can be adapted to sample and trap beyond the 200nm range. Review of the most likely places for ion loss in the sampling process suggests that the sampling and trapping efficiencies can be improved well beyond the 4–5% shown. Moreover, it suggests that sampling of smaller than 10nm ions could achieve efficiencies in the 10s of percent range thereby suggesting new levels of sensitivity can be achieved for small ions (<200kDa). Finally, demonstration of trapping large numbers of 200nm (3×109Da) ions for on-demand ejection in well collimated temporally discrete ion packets is a prelude to resolved mass analysis in that range.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.