Abstract
Abstract Different configurations of a cosmic veto system have been evaluated for the Radionuclide Aerosol Sampler/Analyzer (RASA). The upgrade has the potential to resolve the issue that U.S. International Monitoring System (IMS) radionuclide stations have difficulty meeting the 140Ba Minimum Detectable Concentration (MDC) certification requirement when the natural background radioactivity is high. It also aims to improve the detection sensitivity of the RASA system for other fission and activation products indicative of a nuclear explosion. This proof of principal utilizes a prototype RASA system located at Pacific Northwest National Laboratory (PNNL), and has demonstrated background reductions by up to 11.3% and improvements in the average MDC by up to 7.7%. Recommendations are made for either a 1 or 3 plate configuration, which improves the 140Ba MDC by 5.6% and 7.2% for a blank filter. Whilst elevated 212Pb activity remains problematic, the cosmic veto system provides a practical solution for sensitivity improvement with little design modification of the RASA. It also has value in improving sensitivity and preventing false-positives at remote U.S. IMS station locations where sample transfer for laboratory analysis is logistically difficult.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.